Insertion Loss versus Wavelength
Performance Test
Agilent
8169A
Polarization Controller, Second Edition
123
Insertion Loss versus
Wavelength
1
Make sure all the connectors you will be using are clean.
2
Set up the hardware as shown in Figure D-1.
• For option #020 _rst splice a Agilent 81102BC patchcord to the pigtail of the Pol-
port
• Make sure that all instruments have warmed up.
• Fix all cables with tape so that they won't move during measurements.
3
Set up the 8153A.
a
Zero the 8153A: press
.
b
Set 8153A to dBm : press
until display shows dBm.
c
Set 8153A to 1 sec measuring time: press
until T is shown, set T to
200ms by using the Modify keys.
d
Set 8153A to wavelength=1470 nm: press
until
λ
is shown, set
λ
to 1470 by using the Modify keys.
4
Set up the 8168C.
a
Press
, press
, type 1470.000 on the numeric keys,
press
.
b Press
, press
, type 100 on the numeric keys, press
.
c
Activate the 8168A.
5
Set up the 8169A.
a
Set 8169A to Pol=0 (home position): press
.
b
Select the “Circle" application: press
.
6
Optimize transmission through 8169A:
a
Set the polarizing _lter for maximum transmission (see \Setting the Position
of the Polarizing Filter" in Chapter 2).
b
On 8169A set to 2
ε
Β
and get maximum displayed power on 8153A by turning
knob.
c
On 8169A set to 2
θ
P
and get maximum displayed power on 8153A by turning
knob.