Error Messages
162
Agilent
8169A
Polarization Controller, Second Edition
Display Messages
Selftest Errornnnn. shows that the self test has failed. The number nnnn is
a four digit hexadecimal number that shows which part of the self test has
failed.
So Selftest Error 0010 would mean that the DSP (Digital Signal Processor)
Communications had failed, Selftest Error 12 would mean that the DSP
Communications had failed, and so had the Battery RAM. A value of zero
shows no errors.
Table 31
Bits
Mnemonics
Hexadecimal Value
14
Motor 3
8000
16
13
Motor 2
4000
16
12
Motor 1
1000
16
10
Counter 3
0400
16
9
Counter 2
0200
16
8
Counter 1
0100
16
5
DSP Timeout
0020
16
4
DSP Communications
0010
16
3
Calibration Data
0008
16
1
Battery RAM
0002
16
0
Calibration Data Checksum
0001
16