Performance Test
Insertion Loss versus Wavelength
124
Agilent
8169A
Polarization Controller, Second Edition
Repeat these two steps until absolute maximum on 8153A display is
reached
7
Note the displayed value on 8153A in your test record in the column “Power after
DUT" for the associated wavelength.
8
Repeat list item 6 to list item 7 for the wavelengths between 1480 to 1570 nm in
steps of 10 nm, always setting the 8168A and the 8153A to the required
wavelength for each setting.
9
Repeat list item 6 to list item 7 again for 1470 nm to ensure stability of
measurement setup. If your measured value is more than .01dB off the previous
value, you need to fix your setup and repeat list item 3 to list item 8.
10
Connect the 8168A's output to the 81524A Optical Head as shown in Figure 26
Figure 26 Test Setup for Measuring the Reference PowerIf you're testing an option #020
you need to cut the spliced patchcords first
11
Set 8168A to the wavelength to 1470nm, Output Power to 100
µ
W.
As the display shows negative values, the maximum displayed power is
the smallest number displayed
N O T E