General Cleaning Procedure
Cleaning Information
Agilent
8169A
Polarization Controller, Second Edition
151
To assess the projection of the emitted light beam you can use an infrared
sensor card. Hold the card approximately 5 cm from the interface. The
invisible emitted light is projected onto the card and becomes visible as a
small circular spot.
Optical single mode connections for high-power applications (optical
power levels above 100mW) require careful cleaning to prevent the power
density of burning dust or dirt into the fiber causing permanent damage of
the devices and/or connectors. If this happens, measurements become
inaccurate and non-repeatable.
Preferred Procedure
Use the following procedure on most occasions.
1
Clean the interface by rubbing a new, dry cotton swab over the surface using a
small circular movement.
2
Blow away any remaining lint with compressed air.
Procedure for Stubborn Dirt
Use this procedure when there is greasy dirt on the interface:
1
Moisten a new cotton swab with isopropyl alcohol.
2
Clean the interface by rubbing the cotton swab over the surface using a small
circular movement.
3
Take a new, dry soft tissue and remove the alcohol, dissolved sediment and dust,
by rubbing gently over the surface using a small circular movement.
4
Blow away any remaining lint with compressed air.
This procedures must be strictly followed for each part of the optical connection (connector,
connector interface and physical connector interface).
The connection should be made immediately after cleaning and inspection to prevent the
connection (connector, connector interface and physical connector interface) from getting
dusty or dirty again.