Analyzing the Results
Scanning the Poincare Sphere
Agilent
8169A
Polarization Controller, Second Edition
37
Setting Up the Instruments
1
Run the sphere application with a fast scan.
a Press
b
Make sure that Speed is set to Fast. If it is not, then
• i. Move the Modify knob to start the parameter selection.
• ii. Select Fast using the Modify knob,
or
.
• iii. Press
.
2
Set the averaging time on the power meter to 1s [Press
to select T,
press
to increase T to 1s].
3
Set up a Stability measurement over 20 seconds [Press
, and
to select STABILTY. Press
to select T TOTAL, and set it to 00:00:20, set
AUTODUMP to OFF]
Running the Scan
1
Press
on the polarization controller.
There is a slight delay while the application is initialized, and then the
values of
λ
/4 and
λ
/2 on the display begin to change.
2
When the application is running, read the value for the response of the DUT to a
depolarised signal from the display for the power sensor.
Analyzing the Results
1
When the recording is finished look at the results and find the difference
between the highest and lowest [Press
to get SHOW, press
, and
then
twice to get DIFF].
This is the Polarization Dependent Loss for the DUT.