Error Messages
166
Agilent
8169A
Polarization Controller, Second Edition
Device-Specific Errors
These are error messages in the range -300 to -399, or between 1 and
32767. They show that an error has been detected that is specific to the
operation of the polarization controller.
A device-specific error is signaled by the device-speci_c error bit (bit 3) in
the event status register.
•
300 Device-specific error. This shows that a device-specific error has
occurred. No more specific information is available.
•
310 System error. An instrument system error has occurred.
•
311 Memory error. A memory error has been detected.
•
314 Save/recall memory lost. The nonvolatile data saved by the *SAV
command has been lost.
•
315 Configuration memory lost. The nonvolatile configuration data
saved by the instrument has been lost.
•
330 Self-test failed. Further information about the self-test failure is
available by using *TST?.
•
350 Queue overow. The error queue has overown. This error is written
to the last position in the queue, no further errors are recorded.