General Cleaning Procedure
Cleaning Information
Agilent
8169A
Polarization Controller, Second Edition
155
Preferred Procedure
Use the following procedure on most occasions.
1
Use compressed air at a distance and with low pressure to remove any dust or
lint.
Procedure for Stubborn Dirt
Do not use an ultrasonic bath as this can damage your device.
Use this procedure when there is greasy dirt on the device:
1
Put the optical device into a bath of isopropyl alcohol, and wait at least 10
minutes.
2
Remove the fluid using compressed air at some distance and with low pressure.
If there are any streaks or drying stains on the surface, repeat the whole cleaning
procedure.