Set the Polarizing Filter.
Setting a state of Polarization
Agilent
8169A
Polarization Controller, Second Edition
25
Set the Polarizing Filter.
1
Press
on the polarization controller.
2
Press
.
3
Set the angle of the polarizing _lter for maximum throughput.
a
Type in 10 and press
b
Press
twice to select the tens digit.
c
Using the Modify knob, increase the angle slowly until the power read on the
multimeter increases and then starts to decrease.
d
Press
once to select the units digit.
e
Using the Modify knob, decrease the angle slowly until the power read on the
multimeter starts to decrease.
f
Press
twice to select the hundredths digit.
Using the Modify knob, increase the angle slowly until the power on the
multimeter starts to decrease. Return to the angle that gave the maximum
power.