Installation
Agilent
8169A
Polarization Controller, Second Edition
103
Humidity
The operating humidity for the Agilent 8169A is 15% to 95% from 0°C to
40°C.
Altitude
The Agilent 8169A may operate at up to 10,000 feet.
Installation Category and Pollution Degree
The Agilent 8169A has Installation Category II and Pollution Degree 2
according to IEC 60664.
Instrument Positioning and Cooling
Mount or position the instrument upright and horizontally so that air can
circulate around it freely. When operating the polarization controller,
choose a location that provides at least 75mm (3inches) of clearance at
the rear, and at least 25mm (1inch) of clearance at each side. Failure to
provide adequate air clearance may result in excessive internal temper
ature, reducing instrument reliability.
Figure 21 Correct Positioning of the Polarization Controller