Positioning the Polarizing Filter
Remote Commands
Agilent
8169A
Polarization Controller, Second Edition
67
Positioning the Polarizing Filter
These are the commands that deal with the position of the polarizing filter.
[:INPut]:POSition:POLarizer
Syntax
[:INPut]:POSition:POLarizer <wsp> <value>jMINimumjMAXimumjDEFault
where value is a oating point number between -360.00 and 360.00.
Description
This command sets the position of the polarizing filter. The parameter may
be either
• a number, in mechanical degrees (do not give a unit; the number will be rounded
to the nearest 0.05_),
• MINimum (-360.00°),
• MAXimum (360.00°), or
• DEFault (0.00°).
[:INPut]:POSition:POLarizer?
Syntax
[:INPut]:POSition:POLarizer?
Description
This query gets the position of the polarizing filter in mechanical degrees
(without a unit).
Example
OUTPUT 724;"POS:POL 127"
OUTPUT 724;"POS:POL?"
ENTER 724;A$