General Cleaning Procedure
Cleaning Information
Agilent
8169A
Polarization Controller, Second Edition
147
3
Take a new, dry soft tissue and remove the alcohol, dissolved sediment and dust,
by rubbing gently over the surface using a small circular movement.
4
Blow away any remaining lint with compressed air.
How to clean connector interfaces
Preferred Procedure
Use the following procedure on most occasions.
1
Clean the interface, when no lens is connected, by pushing and pulling a new,
dry pipe cleaner into the opening. Rotate the pipe cleaner slowly as you do this.
2
Blow away any remaining lint with compressed air.
Procedure for Stubborn Dirt
Use this procedure when there is greasy dirt on the interface:
1
Moisten a new pipe cleaner with isopropyl alcohol.
2
Clean the interface by pushing and pulling the pipe cleaner into the opening.
Rotate the pipe cleaner slowly as you do this.
3
Using a new, dry pipe cleaner, and a new, dry cotton swab remove the alcohol,
any dissolved sediment and dust.
4
Blow away any remaining lint with compressed air.
How to clean bare fiber adapters
Bare fiber adapters are difficult to clean. Protect from dust unless they are
in use.
Be careful when using pipe cleaners, as the core and the bristles of the pipe cleaner are
hard and can damage the interface.
Do not use pipe cleaners on optical head adapters, as the hard core of normal pipe cleaners
can damage the bottom of an adapter.
C A U T I O N
Never use any kind of solvent when cleaning a bare fiber adapter as
solvents can:
- Damage the foam inside some adapters.
- Deposit dissolved dirt in the groove, which can then dirty the
surface of an inserted fiber.
C A U T I O N