Connection to a Printer
To connect to a printer you must use a DTE to DCE (
Data Communications
Equipment
) cable.
Set the meter to the TALK ONLY mode by setting the HP-IB address to 31
from the front panel. Then set the RS-232 interface as the active interface.
DTR / DSR Handshake Protocol
The meter is configured as a DTE (
Data Terminal Equipment
) device and
uses the DTR (
Data Terminal Ready
) and DSR (
Data Set Ready
) lines of
the RS-232 interface to handshake. The meter uses the DTR line to send
a hold-off signal. The DTR line must be TRUE before the meter will
accept data from the interface. When the meter sets the DTR line
FALSE, the data must cease within 10 characters.
To disable the DTR/DSR handshake,
do not
connect the DTR line and tie
the DSR line to logic TRUE. If you disable the DTR/DSR handshake, also
select a slower baud rate (300, 600, or 1200 baud) to ensure that the data
is transmitted correctly.
4
Chapter 4 Remote Interface Reference
RS-232 Interface Configuration
179
GPIB
Summary of Contents for 34420A
Page 1: ......
Page 2: ......
Page 13: ...1 Quick Start ...
Page 31: ...2 Front Panel Operation ...
Page 55: ...3 Features and Functions ...
Page 117: ...4 Remote Interface Reference ...
Page 200: ...5 Error Messages ...
Page 213: ...6 Application Programs ...
Page 245: ...6 Chapter 6 Application Programs Microsoft Excel Macro Example 245 ...
Page 246: ...Chapter 6 Application Programs Microsoft Excel Macro Example 246 ...
Page 247: ...7 Measurement Tutorial ...
Page 274: ...8 Specifications ...
Page 279: ...Dimensions Chapter 8 Specifications 34420A Nano Volt Micro Ohm Meter 280 ...
Page 294: ......