To Trigger the Meter
You can trigger the meter from the front panel using
single trigger
or
autotrigger
.
•
Autotriggering is enabled when you turn on the meter. Notice that the
* (sample) annunciator turns on during each measurement.
•
Single triggering takes one reading each time you press
Single
and
then waits for the next trigger. Continue pressing this key to trigger
the meter.
Using an External Trigger
The external trigger mode is also enabled by pressing
Single
.
It is like the single trigger mode except that you apply a trigger pulse
to the rear-panel
Ext Trig
terminal. The meter is triggered on the
negative edge of a TTL pulse.
The front-panel
Single
key is disabled when in remote.
Enables single trigger
and triggers the meter
Toggles between autotrigger
and reading hold
*
(sample) annunciator is on
during each measurement
Trig
annunciator is on when the meter
is waiting for single
trigger (autotrigger disabled)
2
Chapter 2 Front-Panel Operation
To Trigger the Meter
47
Summary of Contents for 34420A
Page 1: ......
Page 2: ......
Page 13: ...1 Quick Start ...
Page 31: ...2 Front Panel Operation ...
Page 55: ...3 Features and Functions ...
Page 117: ...4 Remote Interface Reference ...
Page 200: ...5 Error Messages ...
Page 213: ...6 Application Programs ...
Page 245: ...6 Chapter 6 Application Programs Microsoft Excel Macro Example 245 ...
Page 246: ...Chapter 6 Application Programs Microsoft Excel Macro Example 246 ...
Page 247: ...7 Measurement Tutorial ...
Page 274: ...8 Specifications ...
Page 279: ...Dimensions Chapter 8 Specifications 34420A Nano Volt Micro Ohm Meter 280 ...
Page 294: ......