Calibration Errors
The following errors indicate failures that may occur during a calibration.
Refer to the
Service Guide
for more information.
701
Cal security disabled by jumper
The calibration security feature has been disabled with a jumper inside
the meter. When applicable, this error will occur at power-on to warn you
that the meter is unsecured.
702
Cal secured
The meter is secured against calibration.
703
Invalid secure code
An invalid calibration security code was received when attempting to
unsecure or secure the meter. You must use the same security code to
unsecure the meter as was used to secure it, and vice versa. The security
code may contain up to 11 alphanumeric characters. The first character
must be
a letter.
704
Secure code too long
A security code was received which contained more than 11 characters.
705
Cal aborted
A calibration in progress is aborted when you press any front-panel key,
send a device clear, or change the local/remote state of the meter.
706
Cal value out of range
The specified calibration value (
CAL:VALue
) is invalid for the present
function and range.
707
Cal signal measurement out of range
The specified calibration value (
CAL:VALue
) does not match the signal
applied to the meter.
709
No cal for this function or range
Attempt to calibrate gain on 1 mV range.
710
Full scale correction out of range
Chapter 5 Error Messages
Calibration Errors
210
Summary of Contents for 34420A
Page 1: ......
Page 2: ......
Page 13: ...1 Quick Start ...
Page 31: ...2 Front Panel Operation ...
Page 55: ...3 Features and Functions ...
Page 117: ...4 Remote Interface Reference ...
Page 200: ...5 Error Messages ...
Page 213: ...6 Application Programs ...
Page 245: ...6 Chapter 6 Application Programs Microsoft Excel Macro Example 245 ...
Page 246: ...Chapter 6 Application Programs Microsoft Excel Macro Example 246 ...
Page 247: ...7 Measurement Tutorial ...
Page 274: ...8 Specifications ...
Page 279: ...Dimensions Chapter 8 Specifications 34420A Nano Volt Micro Ohm Meter 280 ...
Page 294: ......