750
Invalid channel name
The meter received a MEAS or CONF command with an invalid channel
specifier.
810
Invalid or unsupported transducer type
An invalid transducer type was detected as a parameter.
820
Temperature out of range for specified transducer
1000
Settings conflict; voltage limit mode has been disabled
After being set to the voltage limited resistance mode, the meter received
a command that changed the settings to an mode or range not allowed.
1001
Settings conflict; power limit mode has been disabled
After being set to the low power resistance mode, the meter received a
command that changed the settings to an mode or range not allowed. For
example, after the low power mode has been enabled, a MEAS:RES
(2-wire Ohms) command is received.
5
Chapter 5 Error Messages
Execution Errors