[{SENSe1: | SENSe2:}]VOLTage[:DC]:RANGe:AUTO {OFF | ON}
[SENSe:]
FRESistance | RESistance
:RANGe:AUTO {OFF | ON}
This command disables or enables autoranging for the function.
Autorange thresholds: Down range at <10% of range; Up range at >120%
of range. [volatile memory]
[{SENSe1: | SENSe2:}]VOLTage[:DC]:RANGe:AUTO?
[SENSe:]
FRESistance | RESistance
:RANGe:AUTO?
This command queries the autorange setting for the specified function.
Returns “0” (
OFF
) or “1” (
ON
).
[SENSe1: | SENSe2:]VOLTage:[DC:]RESolution {<
resolution> | MIN | MAX}
[SENSe:]
FRESistance | RESistance
:RESolution {<
resolution> | MIN | MAX}
This command sets the resolution for the specified function. Specify the
resolution in the same units as the measurement function,
not in number
of digits
. MIN selects the smallest value accepted, which gives the most
resolution. MAX selects the largest value accepted which gives the least
resolution. [volatile memory]
[SENSe1: | SENSe2:]VOLTage:[DC:]RESolution? [{MIN | MAX}]
[SENSe:]
FRESistance | RESistance
:RESolution? [{MIN | MAX}]
This command queries the resolution for the selected function.
Chapter 4 Remote Interface Reference
Setting the Function, Range, and Resolution
138
Summary of Contents for 34420A
Page 1: ......
Page 2: ......
Page 13: ...1 Quick Start ...
Page 31: ...2 Front Panel Operation ...
Page 55: ...3 Features and Functions ...
Page 117: ...4 Remote Interface Reference ...
Page 200: ...5 Error Messages ...
Page 213: ...6 Application Programs ...
Page 245: ...6 Chapter 6 Application Programs Microsoft Excel Macro Example 245 ...
Page 246: ...Chapter 6 Application Programs Microsoft Excel Macro Example 246 ...
Page 247: ...7 Measurement Tutorial ...
Page 274: ...8 Specifications ...
Page 279: ...Dimensions Chapter 8 Specifications 34420A Nano Volt Micro Ohm Meter 280 ...
Page 294: ......