Defaults, Power-On and Reset States
The meter stores settings in either
volatile
or
non-volatile
memory.
Settings stored in
volatile
memory are returned to default settings at
power-on or after a remote reset. Settings stored in
non-volatile
memory
are not changed by power-on or a remote reset (*RST), CONFigure, or
MEASure command.
•
Volatile
memory settings and their defaults:
Setting
Front Panel Default
Remote Default
1
Function
DCV
DCV
Input Channel
Channel 1
Channel 1
Digital Filter
On (50 rdgs)
Off
2
Analog Filter
Off
Off
Offset Compensation
On
Off
N Samples
1
1
Display (On or Off)
On
On
Trigger
Autotrigger
Autotrigger
Math
Off
Off
Readings Hold
Off
Off
Readings Store
Off
Off
Trigger Delay
Auto
Auto
Integration Time (NPLC)
3
(10 plc)
(10 plc)
Null
3
Off
Off
Range
3
Autorange
Autorange
1
Remote defaults are used following reset (*RST), preset (SYSTem:PRESet), CONFigure,
or MEASure commands.
2
You should ensure the digital filter is turned off when using the remote interface. See page 57.
3
Integration Time, Null, and Range are independent in each function. CONFigure or MEASure
commands will only affect the Integration Time, Null, or Range for the function used.
Chapter 3 Features and Functions
Defaults, Power-On and Reset States
112
Summary of Contents for 34420A
Page 1: ......
Page 2: ......
Page 13: ...1 Quick Start ...
Page 31: ...2 Front Panel Operation ...
Page 55: ...3 Features and Functions ...
Page 117: ...4 Remote Interface Reference ...
Page 200: ...5 Error Messages ...
Page 213: ...6 Application Programs ...
Page 245: ...6 Chapter 6 Application Programs Microsoft Excel Macro Example 245 ...
Page 246: ...Chapter 6 Application Programs Microsoft Excel Macro Example 246 ...
Page 247: ...7 Measurement Tutorial ...
Page 274: ...8 Specifications ...
Page 279: ...Dimensions Chapter 8 Specifications 34420A Nano Volt Micro Ohm Meter 280 ...
Page 294: ......