MEASure?
Example
The following program segment shows how to use the
MEASure?
command to make a measurement. This example configures the meter for
voltage measurements on channel 1 using autoranging on the input
signal, automatically places the meter in the “wait-for-trigger” state,
internally triggers the meter to take one reading, and then sends the
reading to the output buffer.
MEAS:VOLT:DC? AUTO,MIN,(@FRONt1)
bus enter statement
This is the simplest way to take a reading. However, you do not have any
flexibility with
MEASure?
to set the trigger count, sample count, trigger
delay, etc. All measurement parameters except function, range, and
resolution are preset for you automatically (
see the table on page 127
CONFigure
Example 1
The following program segment shows how to use the
READ?
command
with
CONFigure
to make an externally-triggered measurement. The
program configures the meter for dc voltage measurements.
CONFigure
does not place the meter in the “wait-for-trigger” state. The
READ?
command places the meter in the “wait-for-trigger” state, takes a reading
when the
Ext Trig
terminal is pulsed, and sends the reading to the
output buffer.
CONF:VOLT:DC 10, MIN, (@FRONt1)
TRIG:SOUR EXT
READ?
bus enter statement
Chapter 4 Remote Interface Reference
Simplified Programming Sequence
132
Summary of Contents for 34420A
Page 1: ......
Page 2: ......
Page 13: ...1 Quick Start ...
Page 31: ...2 Front Panel Operation ...
Page 55: ...3 Features and Functions ...
Page 117: ...4 Remote Interface Reference ...
Page 200: ...5 Error Messages ...
Page 213: ...6 Application Programs ...
Page 245: ...6 Chapter 6 Application Programs Microsoft Excel Macro Example 245 ...
Page 246: ...Chapter 6 Application Programs Microsoft Excel Macro Example 246 ...
Page 247: ...7 Measurement Tutorial ...
Page 274: ...8 Specifications ...
Page 279: ...Dimensions Chapter 8 Specifications 34420A Nano Volt Micro Ohm Meter 280 ...
Page 294: ......