-141
Invalid character data
-144
Character data too long
-148
Character data not allowed
A discrete parameter was received but a character string or a numeric
parameter was expected. Check the list of parameters to verify that you
have used a valid parameter type. Example:
DISP:TEXT ON
-151
Invalid string data
An invalid character string was received. Check to see if you have
enclosed the character string in single or double quotes. Example:
DISP:TEXT ’ON
(the ending quote is missing).
-158
String data not allowed
A character string was received but is not allowed for the command.
Check the list of parameters to verify that you have used a valid
parameter type. Example:
CALC:STAT ’ON’
-160 to -168
Block data errors
The meter does not accept block data.
-170 to -178
Expression errors
An incorrect expression for the channel specifier was received. The meter
does not accept mathematical expressions.
-211
Trigger ignored
A Group Execute Trigger
(
GET
) or
*TRG
was received but the trigger was
ignored. Make sure the meter is in the “wait-for-trigger” state before
issuing a trigger, and make sure the correct trigger source is selected.
-213
Init ignored
An
INITiate
command was received but could not be executed because a
measurement was already in progress. Send a device clear to halt a
measurement in progress and place the meter in the “idle” state.
-214
Trigger deadlock
A trigger deadlock occurs when the trigger source is
BUS
and a
READ?
command is received.
5
Chapter 5 Error Messages
Execution Errors
203
Summary of Contents for 34420A
Page 1: ......
Page 2: ......
Page 13: ...1 Quick Start ...
Page 31: ...2 Front Panel Operation ...
Page 55: ...3 Features and Functions ...
Page 117: ...4 Remote Interface Reference ...
Page 200: ...5 Error Messages ...
Page 213: ...6 Application Programs ...
Page 245: ...6 Chapter 6 Application Programs Microsoft Excel Macro Example 245 ...
Page 246: ...Chapter 6 Application Programs Microsoft Excel Macro Example 246 ...
Page 247: ...7 Measurement Tutorial ...
Page 274: ...8 Specifications ...
Page 279: ...Dimensions Chapter 8 Specifications 34420A Nano Volt Micro Ohm Meter 280 ...
Page 294: ......