Statistics operation commands (AVERage)
•
You must have set
CALC:FUNC AVER
and
CALC ON
to use these
commands.
CALCulate:AVERage:MINimum?
This command returns the minimum value found during a math average
operation. The meter clears the value when math is turned on, when
power has been off, or after a remote interface reset. [volatile memory]
CALCulate:AVERage:MAXimum?
This command returns the maximum value found during a math average
operation. The meter clears the value when math is turned on, when
power has been off, or after a remote interface reset. [volatile memory]
CALCulate:AVERage:AVERage?
This command returns the average of all readings taken since math
average was enabled. The meter clears the value when math is turned
on, when power has been off, or after a remote interface reset.
[volatile memory]
CALCulate:AVERage:COUNt?
This command returns the number of readings taken since math average
was enabled. The meter clears the value when math is turned on, when
power has been off, or after a remote interface reset. [volatile memory]
CALCulate:AVERage:PTPeak?
This command returns the peak-to-peak value of all readings taken since
math was enabled. The meter clears the value when math is turned on,
when power has been off, or after a remote interface reset.
[volatile memory]
CALulate:AVERage:SDEViation
This command returns the standard deviation of all readings taken since
math was enabled. The meter clears the value when math is turned on,
when power has been off, or after a remote interface reset.
[volatile memory]
4
Chapter 4 Remote Interface Reference
Math Operation Commands
149
Summary of Contents for 34420A
Page 1: ......
Page 2: ......
Page 13: ...1 Quick Start ...
Page 31: ...2 Front Panel Operation ...
Page 55: ...3 Features and Functions ...
Page 117: ...4 Remote Interface Reference ...
Page 200: ...5 Error Messages ...
Page 213: ...6 Application Programs ...
Page 245: ...6 Chapter 6 Application Programs Microsoft Excel Macro Example 245 ...
Page 246: ...Chapter 6 Application Programs Microsoft Excel Macro Example 246 ...
Page 247: ...7 Measurement Tutorial ...
Page 274: ...8 Specifications ...
Page 279: ...Dimensions Chapter 8 Specifications 34420A Nano Volt Micro Ohm Meter 280 ...
Page 294: ......