To Set the Parity
You can select the parity for
RS-232
operation. The meter is configured for
even parity with 7 data bits when shipped from the factory.
See also
“Parity Selection,” on page 104.
On/Off
Shift
<
1
Turn on the front-panel menu.
A: MEAS MENU
<
<
2
Move across to the
I/O MENU
choice on this level.
F: I/O MENU
∨
< <
3
Move down a level and then across to the
PARITY
command.
4: PARITY
∨
4
Move down to the “parameter” level to select the parity.
Use the left/right arrow keys to see the parity choices. Choose from one of
the following: None (8 data bits),
Even
(7 data bits), or Odd (7 data bits).
When you set parity, you are indirectly setting the number of data bits.
EVEN: 7 BITS
Auto/Man
5
Save the change and turn off the menu.
ENTER
The parity selection is stored in
non-volatile
memory, and
does not
change when power has been off or after a remote interface reset.
Remote Interface:
you can set or query the parity using the following
commands:
SYStem:COMMunicate:SERial
:PARity {EVEN | ODD | NONE}
:PARity?
4
Chapter 4 Remote Interface Reference
To Set the Parity
191
Summary of Contents for 34420A
Page 1: ......
Page 2: ......
Page 13: ...1 Quick Start ...
Page 31: ...2 Front Panel Operation ...
Page 55: ...3 Features and Functions ...
Page 117: ...4 Remote Interface Reference ...
Page 200: ...5 Error Messages ...
Page 213: ...6 Application Programs ...
Page 245: ...6 Chapter 6 Application Programs Microsoft Excel Macro Example 245 ...
Page 246: ...Chapter 6 Application Programs Microsoft Excel Macro Example 246 ...
Page 247: ...7 Measurement Tutorial ...
Page 274: ...8 Specifications ...
Page 279: ...Dimensions Chapter 8 Specifications 34420A Nano Volt Micro Ohm Meter 280 ...
Page 294: ......