Number of Digits Displayed
Front Panel operation only. See also “Integration Time,” on page 58.
•
You can set the number of digits shown in the display to 4, 5, 6, or 7
full digits
, plus a “
1
⁄
2
” digit which can only be a “0” or “1”.
•
You can set both the number of digits displayed and the Integration
Time in a single operation by pressing digits keys (
•
You can change the number of digits displayed without disturbing
the Integration Time settings by pressing the
<
or
>
keys
(
see page
43).
•
The number of digits is set to 6
1
⁄
2
digits at power-on and after either a
remote interface reset or preset.
•
The number of digits shown is dependent upon integration time and
filter settings. The meter will not allow you to show more digits than
the practical measurement capability of the meter. You can, however,
reduce the number of digits shown. The minimum number of digits
available is 4
1
⁄
2
.
(See the table on page 58)
Chapter 3 Features and Functions
General Measurement Configuration
62
Summary of Contents for 34420A
Page 1: ......
Page 2: ......
Page 13: ...1 Quick Start ...
Page 31: ...2 Front Panel Operation ...
Page 55: ...3 Features and Functions ...
Page 117: ...4 Remote Interface Reference ...
Page 200: ...5 Error Messages ...
Page 213: ...6 Application Programs ...
Page 245: ...6 Chapter 6 Application Programs Microsoft Excel Macro Example 245 ...
Page 246: ...Chapter 6 Application Programs Microsoft Excel Macro Example 246 ...
Page 247: ...7 Measurement Tutorial ...
Page 274: ...8 Specifications ...
Page 279: ...Dimensions Chapter 8 Specifications 34420A Nano Volt Micro Ohm Meter 280 ...
Page 294: ......