Math Operations
There are two math operations available, only one of which can be
enabled at a time. You can choose to either scale the readings as they are
taken, or keep statistics on a group of readings. The selected math
operation remains in effect until you disable it, change functions, turn off
the power, or perform a remote interface reset.
Scaling
•
The scaling is applied as:
reading
=
gain
∗
(
measurement
−
offset
)
Scale
Shift
Stats
•
Front Panel operation:
Set the values of
gain
and
offset
in the
MATH menu. Enable scaling by pressing the scale key. The Stats
annunciator will light in the display.
•
Remote operation:
the math operations and registers are controlled
using commands within the CALCulate command subsystem. First,
select the math operation you want to use
CALCulate:FUNCtion SCALe
Then, set the values for
gain
and
offset
.
CALCulate:SCALe:GAIN <
gain
>
CALCulate:SCALe:OFFSet <
offset
>
Finally, enable the selected math function by turning the math state on:
CALCulate:STATe ON
Chapter 3 Features and Functions
Math Operations
74
Summary of Contents for 34420A
Page 1: ......
Page 2: ......
Page 13: ...1 Quick Start ...
Page 31: ...2 Front Panel Operation ...
Page 55: ...3 Features and Functions ...
Page 117: ...4 Remote Interface Reference ...
Page 200: ...5 Error Messages ...
Page 213: ...6 Application Programs ...
Page 245: ...6 Chapter 6 Application Programs Microsoft Excel Macro Example 245 ...
Page 246: ...Chapter 6 Application Programs Microsoft Excel Macro Example 246 ...
Page 247: ...7 Measurement Tutorial ...
Page 274: ...8 Specifications ...
Page 279: ...Dimensions Chapter 8 Specifications 34420A Nano Volt Micro Ohm Meter 280 ...
Page 294: ......