CONFigure[:VOLTage][:DC]:DIFFerence
[{<
range> | AUTO | MIN | MAX | DEF}][,{<resolution> | MIN | MAX | DEF}]
This command presets and configures the meter for difference
measurements with the specified range and resolution. This command
does not
initiate the measurement. For difference measurements, the
specified range applies to the signal connected to the
Channel 1
terminals. Autoranging is selected for voltage measurements on the
Channel 2
terminals.
CONFigure:FRESistance
CONFigure:RESistance
[{<
range> | AUTO | MIN | MAX | DEF}][,{<resolution> | MIN | MAX | DEF}]
This command presets and configures the meter for 2-wire or 4-wire
ohms measurements with the specified range and resolution. This
command
does not
initiate the measurement.
CONFigure:TEMPerature
{TC | DEF},{B | E | J | K | N | R | S | T | DEF}[,1,{<
resolution> | MIN | MAX | DEF}]
{THER | DEF},{DEF}[,1,{<
resolution> | MIN | MAX | DEF}]
{FRTD | DEF},{85 | 91 | DEF}[,1,{<
resolution> | MIN | MAX | DEF}]
This command presets and configures the meter for temperature
measurements using the specified probe type and resolution. This
command
does not
initiate the measurement. DEF defaults to the last
probe type set.
CONFigure?
This command queries the meter’s present configuration and return a
quoted string.
Chapter 4 Remote Interface Reference
The MEASure? and CONFigure Commands
136
Summary of Contents for 34420A
Page 1: ......
Page 2: ......
Page 13: ...1 Quick Start ...
Page 31: ...2 Front Panel Operation ...
Page 55: ...3 Features and Functions ...
Page 117: ...4 Remote Interface Reference ...
Page 200: ...5 Error Messages ...
Page 213: ...6 Application Programs ...
Page 245: ...6 Chapter 6 Application Programs Microsoft Excel Macro Example 245 ...
Page 246: ...Chapter 6 Application Programs Microsoft Excel Macro Example 246 ...
Page 247: ...7 Measurement Tutorial ...
Page 274: ...8 Specifications ...
Page 279: ...Dimensions Chapter 8 Specifications 34420A Nano Volt Micro Ohm Meter 280 ...
Page 294: ......