Remote operation:
The CONFigure and MEASure subsystems use
an optional parameter to specify the input channel as either 1 or 2:
CONF:VOLT:DC DEF, DEF,(@FRONTl)
CONF:VOLT:DC DEF, DEF,(@FRONT2)
Channel 1 operation
Channel 2 operation
The SENSe subsystem uses a keyword modification to indicate the
channel:
SENS1:VOLT:DC:RANG:AUTO
SENS2:VOLT:DC:RANG:AUTO
Channel 1 operation
Channel 2 operation
Two Channel Measurements
Ch1 - Ch2
•
Difference measurements:
First, set the range and null value for
each input channel. Then, select difference measurements.
Difference
=
((
Ch
1
voltage
−
Ch
1
Null
)
−
(
Ch
2
voltage
−
Ch
2
Null
))
−
Null
*
Shift
Ch1- Ch2
•
Ratio measurements:
First, set the range and null value for each
input channel. Then, select ratio measurements.
Ratio
=
(
Ch
1
voltage
−
Ch
1
Null
)
(
Ch
2
voltage
−
Ch
2
Null
)
•
Remote operation:
use the following commands to measure or
configure a difference or ratio measurement. Set each channel to the
appropriate range and null conditions using:
[SENSe1: | SENSe2:]
VOLTage:DC
:RANGE <
range
>
:NULL:STATe {OFF | ON}
:NULL:VALue {<
value
> | MIN | MAX}
Then use the following commands to measure the ratio or difference:
MEASure:VOLTage:DC
:RATIO?
:DIFFerence?
CONFigure:VOLTage:DC
:RATIO
:DIFFerence
[SENSe:]FUNCtion "VOLTage:RATio"
[SENSe:]FUNCtion "VOLTage:DIFFerence"
*Difference Null is only available from the Front Panel.
3
Chapter 3 Features and Functions
Voltage Measurement Configuration
67
Summary of Contents for 34420A
Page 1: ......
Page 2: ......
Page 13: ...1 Quick Start ...
Page 31: ...2 Front Panel Operation ...
Page 55: ...3 Features and Functions ...
Page 117: ...4 Remote Interface Reference ...
Page 200: ...5 Error Messages ...
Page 213: ...6 Application Programs ...
Page 245: ...6 Chapter 6 Application Programs Microsoft Excel Macro Example 245 ...
Page 246: ...Chapter 6 Application Programs Microsoft Excel Macro Example 246 ...
Page 247: ...7 Measurement Tutorial ...
Page 274: ...8 Specifications ...
Page 279: ...Dimensions Chapter 8 Specifications 34420A Nano Volt Micro Ohm Meter 280 ...
Page 294: ......