TRIGger:DELay {<
seconds> | MIN | MAX}
This command inserts a trigger delay between the trigger signal and each
sample that follows. If you do not specify a trigger delay, the meter
automatically selects a delay for you (
see page
89). Select from 0 to 3600
seconds. MIN = 0 seconds. MAX = 3600 seconds. [volatile memory]
TRIGger:DELay? [{MIN | MAX}]
This command queries the trigger delay.
TRIGger:DELay:AUTO {OFF | ON}
This command enables or disables an automatic trigger delay. The delay
is determined by function, range, integration time, and ac filter setting.
Selecting a specific trigger delay value automatically turns off automatic
trigger delay. [volatile memory]
TRIGger:DELay:AUTO?
This command queries the automatic trigger delay setting.
Returns “0” (
OFF
) or “1” (
ON
).
SAMPle:COUNt {<
value> | MIN | MAX}
This command sets the number of readings (samples) the meter takes per
trigger. Select from 1 to 50,000 readings per trigger.
MIN
= 1 and
MAX
= 50,000. [volatile memory]
SAMPle:COUNt? [{MIN | MAX}]
This command queries the sample count.
TRIGger:COUNt {<
value> | MIN | MAX | INFinity}
This command sets the number of triggers the meter will accept before
returning to the “idle” state. Select from 1 to 50,000 triggers. The
INFinity
parameter instructs the meter to continuously accept triggers
(you must send a device clear to return to the “idle” state). Trigger count
is ignored while in local operation. MIN = 1 and MAX = 50,000. [volatile
memory]
TRIGger:COUNt? [{MIN | MAX}]
This command queries the trigger count. If you specify an infinity trigger
count, the query command returns “9.9037”.
4
Chapter 4 Remote Interface Reference
Triggering Commands
155
Summary of Contents for 34420A
Page 1: ......
Page 2: ......
Page 13: ...1 Quick Start ...
Page 31: ...2 Front Panel Operation ...
Page 55: ...3 Features and Functions ...
Page 117: ...4 Remote Interface Reference ...
Page 200: ...5 Error Messages ...
Page 213: ...6 Application Programs ...
Page 245: ...6 Chapter 6 Application Programs Microsoft Excel Macro Example 245 ...
Page 246: ...Chapter 6 Application Programs Microsoft Excel Macro Example 246 ...
Page 247: ...7 Measurement Tutorial ...
Page 274: ...8 Specifications ...
Page 279: ...Dimensions Chapter 8 Specifications 34420A Nano Volt Micro Ohm Meter 280 ...
Page 294: ......