MEASure[:VOLTage][:DC]:DIFFerence?
[{<
range> | AUTO | MIN | MAX | DEF}][,{<resolution> | MIN | MAX | DEF}]
This command presets and makes a difference measurement with the
specified range and resolution. The reading is sent to the output buffer.
For difference measurements, the specified range applies to the signal
connected to the
Channel 1
terminals. Autoranging is selected for voltage
measurements on the
Channel 2
terminals.
MEASure:FRESistance?
MEASure:RESistance?
[{<
range> | AUTO | MIN | MAX | DEF}][,{<resolution> | MIN | MAX | DEF}]
This command presets and makes a 2-wire or 4-wire ohms measurement
with the specified range and resolution. The reading is sent to the output
buffer.
MEASure:TEMPerature?
{TC | DEF},{B | E | J | K | N | R | S | T | DEF}[,1,{<
resolution> | MIN | MAX | DEF}]
{THER | DEF},{DEF}[,1,{<
resolution> | MIN | MAX | DEF}]
{FRTD | DEF},{85 | 91 | DEF}[,1,{<
resolution> | MIN | MAX | DEF}]
This command presets and configures the meter for temperature
measurements using the specified probe type and resolution. The reading
is sent to the output buffer. DEF defaults to the last probe type set.
CONFigure[:VOLTage][:DC]
[{<
range> | AUTO | MIN | MAX | DEF}][,{<resolution> | MIN | MAX | DEF}][,(@<channel>)]
This command presets and configures the meter for dc voltage
measurements with the specified range, resolution, and input channel.
This command
does not
initiate the measurement.
CONFigure[:VOLTage][:DC]:RATio
[{<
range> | AUTO | MIN | MAX | DEF}][,{<resolution> | MIN | MAX | DEF}]
This command presets and configures the meter for dc:dc ratio
measurements with the specified range and resolution. This command
does not
initiate the measurement. For ratio measurements, the specified
range applies to the signal connected to the
Channel 1
terminals.
Autoranging is selected for voltage measurements on the
Channel 2
terminals.
4
Chapter 4 Remote Interface Reference
The MEASure? and CONFigure Commands
135
Summary of Contents for 34420A
Page 1: ......
Page 2: ......
Page 13: ...1 Quick Start ...
Page 31: ...2 Front Panel Operation ...
Page 55: ...3 Features and Functions ...
Page 117: ...4 Remote Interface Reference ...
Page 200: ...5 Error Messages ...
Page 213: ...6 Application Programs ...
Page 245: ...6 Chapter 6 Application Programs Microsoft Excel Macro Example 245 ...
Page 246: ...Chapter 6 Application Programs Microsoft Excel Macro Example 246 ...
Page 247: ...7 Measurement Tutorial ...
Page 274: ...8 Specifications ...
Page 279: ...Dimensions Chapter 8 Specifications 34420A Nano Volt Micro Ohm Meter 280 ...
Page 294: ......