Configuration and Activation 4-89
5.
Continue with other activities that require communication with
the monitor or disconnect from the monitor by clicking on the
green
Hangup
button in the
Functions
section. If applicable
to the connection type, consider checking the monitor status to
verify the monitor is active and the data storage interval
reflects the sample rate selected for the monitor.
Important Note
: When installing a
TRITON
+
monitor
with an AV|Gated, AV|Max, ParaDepth or ParaFlow
sensor, the next step after the monitor activation and
sensor installation is to verify the sensor is installed
properly with a
Sensor Rotation
and
Sensor Slope
of no
more than +/- 5 degrees. The closer the
Rotation
and
Slope
are to zero, the better. Proceed to
Running Sensor
Diagnostics
on page 4-92 for more details.
Q
start
XML
generates a log entry of the activation event which is
available for viewing through the log viewer in
Q
start
XML
.
Summary of Contents for 507165
Page 20: ...1 12 ADS TRITON Manual Control Drawing ...
Page 27: ...Introduction 1 19 CS6 LRD Sensor Certificate of Conformity ...
Page 29: ...Introduction 1 21 CSA Certificate of Compliance Page 1 of 7 ...
Page 30: ...1 22 ADS TRITON Manual CSA Certificate of Compliance Page 2 of 7 ...
Page 31: ...Introduction 1 23 CSA Certificate of Compliance Page 3 of 7 ...
Page 32: ...1 24 ADS TRITON Manual CSA Certificate of Compliance Page 4 of 7 ...
Page 33: ...Introduction 1 25 CSA Certificate of Compliance Page 5 of 7 ...
Page 34: ...1 26 ADS TRITON Manual CSA Certificate of Compliance Page 6 of 7 ...
Page 35: ...Introduction 1 27 CSA Certificate of Compliance Page 7 of 7 ...
Page 37: ...Introduction 1 29 Sample TRITON Declaration of Conformity Page 2 ...
Page 53: ...System Overview 2 13 ExPAC XBUS ...
Page 266: ...5 60 ADS TRITON Manual z Installing the Submersion Sensor z Measuring the physical offset ...
Page 288: ...5 82 ADS TRITON Manual Topside Wall Mount Bracket Topside Sensor Bracket ...
Page 345: ...7 2 ADS TRITON Manual use of this installation guide or the installation of any ADS equipment ...
Page 358: ...Inclinometer 8 3 ...
Page 507: ...C 4 ADS TRITON Manual ...