Maintenance and Troubleshooting 11-59
Problem
Individual ultrasonic depth readings (SDEPTH_3) from
ParaDepth, ParaFlow, Surface Combo
Sensor/Ultrasonic Depth Sensor are consistently
equivalent to the value obtained from the following
equation: Pipe Height
–
Physical 9.99.
Possible
Causes
Sensor may be faulty.
5-volt (blue) fuse on regulator may have blown.
Possible
Solutions
Replace the Surface Combo Sensor/Ultrasonic Depth
Sensor.
Replace the 5-volt (blue) fuse on the regulator.
Caution:
Do not connect a potentially shorted sensor to
another monitor until confirming that the sensor is
functional (i.e., not shorted) using a voltmeter. A shorted
sensor will blow the 5-volt (blue) replacement fuse on
the regulator of the monitor.
Problem
Sensor reports abnormal temperatures.
Possible
Causes
Sensor connection to monitor may be loose.
Temperature sensor may be faulty.
Sensor may be faulty.
Possible
Solutions
Secure sensor connection to monitor if necessary.
Replace the sensor.
Summary of Contents for 507165
Page 20: ...1 12 ADS TRITON Manual Control Drawing ...
Page 27: ...Introduction 1 19 CS6 LRD Sensor Certificate of Conformity ...
Page 29: ...Introduction 1 21 CSA Certificate of Compliance Page 1 of 7 ...
Page 30: ...1 22 ADS TRITON Manual CSA Certificate of Compliance Page 2 of 7 ...
Page 31: ...Introduction 1 23 CSA Certificate of Compliance Page 3 of 7 ...
Page 32: ...1 24 ADS TRITON Manual CSA Certificate of Compliance Page 4 of 7 ...
Page 33: ...Introduction 1 25 CSA Certificate of Compliance Page 5 of 7 ...
Page 34: ...1 26 ADS TRITON Manual CSA Certificate of Compliance Page 6 of 7 ...
Page 35: ...Introduction 1 27 CSA Certificate of Compliance Page 7 of 7 ...
Page 37: ...Introduction 1 29 Sample TRITON Declaration of Conformity Page 2 ...
Page 53: ...System Overview 2 13 ExPAC XBUS ...
Page 266: ...5 60 ADS TRITON Manual z Installing the Submersion Sensor z Measuring the physical offset ...
Page 288: ...5 82 ADS TRITON Manual Topside Wall Mount Bracket Topside Sensor Bracket ...
Page 345: ...7 2 ADS TRITON Manual use of this installation guide or the installation of any ADS equipment ...
Page 358: ...Inclinometer 8 3 ...
Page 507: ...C 4 ADS TRITON Manual ...
Page 535: ...Index In 19 wiring to the monitor 6 29 XIO device parameters 4 84 ...