4-34 ADS
®
Manual
z
Enable
Select this checkbox to ensure the monitor begins
taking velocity readings at the designated interval once
activated. This option is selected by default.
z
Flow Type
Select the option that best reflects the kind of
flow conditions that typically occur at the monitoring point
from the drop-down list:
Typical
Flow exhibits a typical diurnal pattern.
This is
the default setting.
Wet/Stagnant
Flow remains in the pipe, but moves only
during wet weather events.
Usually Dry
Pipe typically remains dry, except under
wet weather conditions.
z
Velocity Direction
Select one of the following options that
reflects the general direction of the flow at the monitoring
point and position of the velocity sensor.
The default is
Forward
.
Bidirectional
Choose this option for monitoring points
that potentially experience reverse flows. The AV|Gated
sensor typically is positioned facing upstream into the
flow. Choosing this ensures the monitor analyzes both
positive and negative values returned from the velocity
sensor.
Forward
Choose this option for monitoring points where
reverse flows are
not
expected and to ensure the monitor
only analyzes positive values returned from the velocity
sensor. The AV|Gated sensor is typically positioned
facing upstream into the flow. This is the most common,
and the default, option for selection. ADS recommends
using this setting for most monitoring applications.
Forward
is the default selection for
Velocity Direction
.
Backward
Choose this option for monitoring points
where the AV|Gated is installed facing downstream
(backwards) from the oncoming flow. In locations where
there is high risk of debris hitting the sensor, this might be
the desired option.
Summary of Contents for 507165
Page 20: ...1 12 ADS TRITON Manual Control Drawing ...
Page 27: ...Introduction 1 19 CS6 LRD Sensor Certificate of Conformity ...
Page 29: ...Introduction 1 21 CSA Certificate of Compliance Page 1 of 7 ...
Page 30: ...1 22 ADS TRITON Manual CSA Certificate of Compliance Page 2 of 7 ...
Page 31: ...Introduction 1 23 CSA Certificate of Compliance Page 3 of 7 ...
Page 32: ...1 24 ADS TRITON Manual CSA Certificate of Compliance Page 4 of 7 ...
Page 33: ...Introduction 1 25 CSA Certificate of Compliance Page 5 of 7 ...
Page 34: ...1 26 ADS TRITON Manual CSA Certificate of Compliance Page 6 of 7 ...
Page 35: ...Introduction 1 27 CSA Certificate of Compliance Page 7 of 7 ...
Page 37: ...Introduction 1 29 Sample TRITON Declaration of Conformity Page 2 ...
Page 53: ...System Overview 2 13 ExPAC XBUS ...
Page 266: ...5 60 ADS TRITON Manual z Installing the Submersion Sensor z Measuring the physical offset ...
Page 288: ...5 82 ADS TRITON Manual Topside Wall Mount Bracket Topside Sensor Bracket ...
Page 345: ...7 2 ADS TRITON Manual use of this installation guide or the installation of any ADS equipment ...
Page 358: ...Inclinometer 8 3 ...
Page 507: ...C 4 ADS TRITON Manual ...
Page 535: ...Index In 19 wiring to the monitor 6 29 XIO device parameters 4 84 ...