Configuration and Activation 4-11
Enter the information for each monitoring point in the following
way:
1.
Select the type or shape of the pipe or channel in which the
sensors are installed from the
Type
drop-down list on the ADS
Q
start
XML
dialog.
2.
Enter the installation information in the corresponding fields
based on the selected pipe or installation type.
Pipe
Height
Enter the distance from the bottom of the
pipe to the crown (i.e., top or peak).
Width
Enter the horizontal distance across the widest
portion of the pipe.
Note:
If the width measurement does not equal the height
measurement, select
Elliptical
for the pipe shape.
Weir
Depth Range
If the ultrasonic depth sensor is
installed upstream from the weir wall, enter the
distance from the face of the sensor to the bottom of
the channel. If the ultrasonic depth sensor is mounted
directly over the weir wall, enter the distance from the
face of the sensor to the crest of the weir.
Const K
This field is read-only for most weir
installations, except
Rectangular
and
Trapezoidal
weirs with
Constant K values
. The monitor uses the
default Constant K value to convert weir flows to CFS
(cubic feet per second) and to calculate quantity.
Entering a different Constant K value will modify the
weir equation, potentially resulting in inaccurate CFS
quantities. Therefore, please verify all manual
modifications to this value.
Capacity
Enter the maximum capacity of the weir
based on the type of weir and any parameters applied
and/or measurements taken at the location for the
application.
Summary of Contents for 507165
Page 20: ...1 12 ADS TRITON Manual Control Drawing ...
Page 27: ...Introduction 1 19 CS6 LRD Sensor Certificate of Conformity ...
Page 29: ...Introduction 1 21 CSA Certificate of Compliance Page 1 of 7 ...
Page 30: ...1 22 ADS TRITON Manual CSA Certificate of Compliance Page 2 of 7 ...
Page 31: ...Introduction 1 23 CSA Certificate of Compliance Page 3 of 7 ...
Page 32: ...1 24 ADS TRITON Manual CSA Certificate of Compliance Page 4 of 7 ...
Page 33: ...Introduction 1 25 CSA Certificate of Compliance Page 5 of 7 ...
Page 34: ...1 26 ADS TRITON Manual CSA Certificate of Compliance Page 6 of 7 ...
Page 35: ...Introduction 1 27 CSA Certificate of Compliance Page 7 of 7 ...
Page 37: ...Introduction 1 29 Sample TRITON Declaration of Conformity Page 2 ...
Page 53: ...System Overview 2 13 ExPAC XBUS ...
Page 266: ...5 60 ADS TRITON Manual z Installing the Submersion Sensor z Measuring the physical offset ...
Page 288: ...5 82 ADS TRITON Manual Topside Wall Mount Bracket Topside Sensor Bracket ...
Page 345: ...7 2 ADS TRITON Manual use of this installation guide or the installation of any ADS equipment ...
Page 358: ...Inclinometer 8 3 ...
Page 507: ...C 4 ADS TRITON Manual ...
Page 535: ...Index In 19 wiring to the monitor 6 29 XIO device parameters 4 84 ...