Sensor Installation and Connection 5-29
3.
Place the ring in the input pipe
at least
12 inches (305 mm)
upstream from the manhole or edge of the pipe with the
sensors facing upstream toward the oncoming flow. It must
be located far enough upstream from the manhole to minimize
the effect of the draw-down caused by a possible drop in the
manhole invert.
Installing the ring at least 12 inches (305 mm) upstream from the manhole
invert
Keep the following in mind:
The installed AV|Gated sensor must be positioned in the
bottom, center of the pipe and
cannot /- 5
degrees from the 6:00 position.
If a greater rotation
/- 5 degrees is needed, then a Silt Mount Adapter
must be used. Install the AV|Gated sensor and then verify
the sensor is correctly positioned. After activating the
monitor, perform a Diagnostic reading on the AV|Gated
Ultrasonic sensor and verify the
Sensor Slope
and
Sensor
Rotation
values are not greater than 5.0. See
Chapter 4,
Running Sensor Diagnostics
for more information.
Be sure to mount the ParaDepth, ParaFlow, Surface
Combo sensor/Ultrasonic Depth Sensor at the top (crown)
of the pipe and to mount the AV|Gated, AV|Max, and Peak
Combo sensors as close as possible to the bottom center of
the pipe,
above
any silt present and
below
the flow surface
(during minimum flows).
FLOW
12"
Summary of Contents for 507165
Page 20: ...1 12 ADS TRITON Manual Control Drawing ...
Page 27: ...Introduction 1 19 CS6 LRD Sensor Certificate of Conformity ...
Page 29: ...Introduction 1 21 CSA Certificate of Compliance Page 1 of 7 ...
Page 30: ...1 22 ADS TRITON Manual CSA Certificate of Compliance Page 2 of 7 ...
Page 31: ...Introduction 1 23 CSA Certificate of Compliance Page 3 of 7 ...
Page 32: ...1 24 ADS TRITON Manual CSA Certificate of Compliance Page 4 of 7 ...
Page 33: ...Introduction 1 25 CSA Certificate of Compliance Page 5 of 7 ...
Page 34: ...1 26 ADS TRITON Manual CSA Certificate of Compliance Page 6 of 7 ...
Page 35: ...Introduction 1 27 CSA Certificate of Compliance Page 7 of 7 ...
Page 37: ...Introduction 1 29 Sample TRITON Declaration of Conformity Page 2 ...
Page 53: ...System Overview 2 13 ExPAC XBUS ...
Page 266: ...5 60 ADS TRITON Manual z Installing the Submersion Sensor z Measuring the physical offset ...
Page 288: ...5 82 ADS TRITON Manual Topside Wall Mount Bracket Topside Sensor Bracket ...
Page 345: ...7 2 ADS TRITON Manual use of this installation guide or the installation of any ADS equipment ...
Page 358: ...Inclinometer 8 3 ...
Page 507: ...C 4 ADS TRITON Manual ...
Page 535: ...Index In 19 wiring to the monitor 6 29 XIO device parameters 4 84 ...