13
Figure 17 NGC Optional Mounting Flange Pipe
5)
If the Optional Mounting Pipe with Flange is used, insert Hex Socket Screw
through hole in the welded flange into the neck bottom of unit and tighten
using ¼” Hex Wrench. Repeat for all screws.
6)
Small adjustments may be made to orientation, by applying additional
pressure to mounting pipe with pipe wrench, tightening mounting pipe into
shelf mounted flange or pipe saddle flange.
NOTE:
When positioning the unit, you should take into consideration the
mounting of the Sample Conditioning System, Conduit locations and
access to the rear end cap of the unit.
IX.
Sample Conditioning Module Installation
Both Sample Conditioning Module Mounting Brackets are installed identically. The
Single Module Bracket will hold a single stream Sample Conditioning Module, while the
Multiple Module Bracket will hold up to 3 Sample Conditioning Modules.
A. Materials
•
Installed
NGC
•
Single or Multiple Module Mounting Kit
•
1 ea. .312 x 2.5 x 3.62 x 1.5 U-Bolt
•
2 ea. 5/16” SST Split Washer
•
2 ea. 5/16” SST Flat Washer
•
2 ea. 5/16-18 SST Lock Nut
•
Sample Conditioning Module(s) and hardware
B. Instructions
1)
On Sample Conditioning Module, align mounting holes to corresponding
holes in bracket. Insert Bolt through hole in bracket, from front to back
through mounting hole in module (see Figure 18). Place split washer, then flat