Pannoramic 250 FLASH III 3.0.2 User’s Guide
5.1.2 Microscope Control Panel
Preview Type
setting determines the type of illumination. In Brightfield mode, the following preview
types are available:
•
Brightfield preview
– a normal type of illumination in the microscope for scanning the
slide. This option is recommended when scanning in Brightfield mode.
Remove Coverslip
– If selected, the edges of the coverslip are not detected as specimen to be
scanned. Depending on the threshold level, it is possible that the edges of the coverslip are detected
as areas to be digitized.
Figure 32 – Preview image with and without coverslip edge
Caution!
It requires lots of time (since it cannot find focus neither on the edges,
nor on specks) to digitize coverslip edges. Select
Remove coverslip
function to eliminate those areas.
Warning
!
Do
NOT
use
Remove coverslip
function on TMA slides! The layout of
TMA slides may mislead the detecting algorithm.