Pannoramic 250 FLASH III 3.0.2 User’s Guide
5.1.2 Microscope Control Panel
Enter the distance between two spots on the X and the Y axis in µm in the
Core distance
fields or select the appropriate values with the arrows. Enter the average diameter of the
spots in the
Core size
field or select the appropriate value with the arrows.
•
Clear scanmap
– The entire scanmap area will be cleared and reset to default.
•
Use selection box restriction
– The preview image shows the area that will be digitized
with a red and/or an orange mask. If the
Use selection box
is enabled, and the crop function
is selected, a black frame displayed in the image, and only the area inside the box is scanned.
It is useful if you only want to scan the area that is equivalent with the size of the coverslip, so
external specks will not be scanned.
Figure 37 – Preview image
The following four colors are used in the preview image to demonstrate digitization and selection
settings:
September 16, 2020 – Rev. 1
3DHISTECH Ltd.
70(156)