![background image](http://html1.mh-extra.com/html/3dhistech/pannoramic-250-flash-iii/pannoramic-250-flash-iii_user-manual_442167133.webp)
Pannoramic 250 FLASH III 3.0.2 User’s Guide
5.6.2 Service
Click
Start slide insert test
button to launch the process. Slides are loaded in but preview
image will not be recorded.
If
Magazine counter
is set to -1, all the magazines will be tested in cycles based on the value
set in the
Loading cycle per magazine
field.
Click
Remove magazine
when finished.
5.7 Hosting
Hosting a live teleconsultation session gives specialists the opportunity to examine the same Live
microscopic image during a real-time counseling even if being in laboratories far away from each
other. This means multiple users can connect to the device at the same time. The user of the scanner
will be the host with direct access to the scanner.
Note:
Both the client's and the host's scanner software version must be
identical.
Once client users have connected, they will be able to reach the following functions of scanner:
•
Live image
•
Preview window
•
Profile content
•
Stage control
Each participant see the same live image within the same preview window. The active user can
navigate on the slide and set up a scanning profile. The
Start scan
function is also available for the
active user. Each participant can make comments in a chat window while examining the slide image.
Warning!
The user who starts hosting should not touch the hardware after the
control has been passed on to avoid personal injury from possible
hardware movements (for example, inserting new cassettes or slides,
installing fluorescent filters, etc.).
September 16, 2020 – Rev. 1
3DHISTECH Ltd.
133(156)