ERROR TITLE
ERROR MESSAGE
CAUSE / ACTIONS
No Pulse Oximeter
Reading
Confirm the Pulse Oximeter is
connected to the Recorder and
correctly positioned on your finger.
The study will resume automatically
after fixing the error.
Pulse Oximeter is not connected to
Reconnect or not positioned properly
on your finger.
•
•
•
Disconnect/Reconnect Pulse Oximeter
to Recorder
Ensure sensor is properly positioned
on your finger
If error continues, stop the study and
contact your Healthcare Provider
No Effort Belt Reading
Confirm the Effort Belt is connected to
the Recorder and correctly positioned
on your abdomen. The study will
resume automatically after fixing the
error.
Effort Belt is not connected to
Recorder, or not positioned correctly
on your abdomen.
•
•
•
•
Disconnect/Reconnect Effort Belt to
Recorder
Disconnect/reconnect Effort Belt cable
to Effort Belt
Ensure Effort Belt is properly
positioned on your abdomen (i.e. in
front of your belly button)
If error continues, stop the study and
contact your Healthcare Provider
Positioner Error
The MP’s motor has stalled; this may
be caused by pushing against the trays
while they are moving in your mouth.
To correct this error, press Continue and
this will pause tray movement for 30
minutes and allow you to fall asleep.
•
•
Press Continue and this will pause tray
movement for 30 minutes and allow
you to fall asleep
If this error continues, the MP may
not be functioning properly. Stop the
study and contact your Healthcare
Provider.
No Further Studies
Return MATRx plus to your Healthcare
Provider
There have been 6 unsuccessful study
nights and there are no more studies
ordered.
• Return MATRx plus to your Healthcare
Provider
No Further Studies
Study is complete. Return MATRx plus
to your Healthcare Provider.
The study has been completed
successfully and is ready for evaluation.
• Return MATRx plus to your Healthcare
Provider
Error
Device error. Return MATRx plus to your
Healthcare Provider.
Maintenance by manufacturer is
required.
• Return MATRx plus to your Healthcare
Provider
Recorder Plugged In
Confirm the Recorder’s power adapter
is not plugged into the Recorder. The
study will resume automatically after
•
The Recorder should not be plugged
in during the study. Disconnect the
power adapter from the Recorder to
continue the study.
68