With very low perfusion at the monitored site, the readings may read lower than core arterial oxygen
saturation
Placement of a sensor on an extremity with a blood pressure cuff, arterial catheter, or intravascular line
Loss of pulse signal can occur in any of the following situations:
13.
The Pulse Oximeter sensor is too tight
There is excessive illumination from light sources such as a surgical lamp, a bilirubin lamp, or sunlight
A blood pressure cuff is inflated on the same extremity as the one with the Pulse Oximeter attached
The patient has hypotension, severe vasoconstriction, severe anemia, or hypothermia
There is arterial occlusion proximal to the Pulse Oximeter
The patient is in cardiac arrest or is in shock
•
•
•
•
•
•