Oxygen Saturation (the instantaneous saturation as well as the
calculated baseline) is displayed as a line graph in percent.
Baseline
- calculated from a 5 minute moving average window
Saturation
- the oxygen saturation at any given time
3.
4.
5.
Nasal Airflow is displayed as a line graph in arbitrary values.
Flow Total
- Combined airflow from both nostrils
L Naris Pressure
- Airflow measured from the left nostril
R Naris Pressure
- Airflow measured from the right nostril
Zero Line
- Zero airflow line
Zero Range
- Airflow is not measured below these lines
Heart Rate is displayed as a line graph in beats per minute (BPM).
Respiratory Effort is displayed as a line graph in arbitrary values.
6.
Sensor Status is displayed as a color bar graph at the top of the
signal display.
Active
(sensors on, study running) -
Green
Inactive
(sensors on, study will resume) -
Orange
Paused
(sensors on, awaiting user input to continue study) -
Red
Excluded
(manually excluded time) -
Black
Alert
(patient was alerted to problem with sensor) -
Exclamation
mark (!)
- Position the mouse over the (!) for the alert type.
Mandibular Protrusion is displayed as a line graph in millimeters
Max
- The maximum protrusion level of the patient
Feedback
- The actual position of the MP at any given time
Control
- The commanded position of the MP at any given time
1.
2.
45
Note
Only data collected during the ‘Active’ Study will be used in the OA
Study analysis
Caution
The control and feedback position of the MP should not differ by more
than 0.5mm. If greater differences are noted, conduct a Functional
Check or contact the Manufacturer.