9-12
DATABASE MANAGEMENT
REV. B 02/03 PN 51680-1
HUMPHREY FIELD ANALYZER
II -
i
series
Hard drive failure:
If you are faced with a hard drive failure, and you elect to attempt recovery immediately,
FIX NOW option
choose the FIX NOW option. This option should only be initiated if you have the time, and
patients will not be inconvenienced. Otherwise, elect to save tests to floppy disk only, and
attempt recovery at a later time (see FLOPPY ONLY option). To help you decide, a screen
message alerts you to the approximate recovery time.
Note: If you were to choose FIX NOW, and the attempt failed, you could always resort to
FLOPPY ONLY, if patients were scheduled for testing. In that case, once patient testing was
completed, you would attempt to rebuild the hard drive database (see REBUILD HARD
DISK DATABASE option).
1
Read the message in the
dialog box. Choose FIX
NOW.
2
While the HFA II is
attempting to recover the
hard disk database, the screen
displays a progress bar represent-
ing processing time completed.
9.htp
3/14/03, 3:19 PM
12
Содержание HUMPHREY 720i
Страница 1: ...HUMPHREY FIELD ANALYZER II i series USER S GUIDE Model 720i Model 740i Model 745i Model 750i...
Страница 4: ......
Страница 26: ...1 18 INTRODUCTION INSTRUMENT SETUP REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Страница 116: ...6 8 TEST RELIABILITY REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Страница 208: ...10 20 CUSTOM TESTING REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Страница 284: ...12 14 CARE AND CLEANING REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Страница 327: ......
Страница 328: ......