11-13
REV. B 02/03 PN 51680-1
HUMPHREY FIELD ANALYZER
II -
i
series
KINETIC TESTING
Unwanted test points can be removed from the Kinetic test you are performing. However, only
the last entry can be removed. If you are unsure of a patient response, it is best to erase it and
test that meridian again.
1
From the Kinetic Test
screen, press DELETE LAST
POINT.
The last entry you have made
will automatically be removed
from the test.
No warning will
be issued before deleting
the last entry.
2
Subsequent selections of the DELETE LAST POINT button will erase the next previous
entry.
Note: The DELETE LAST POINT button is disabled if there are no entries.
You may wish to retest a previously tested meridian. If you have not deleted the previous
response, characters representing both responses will appear on the display and in the kinetic
results printout. However, only the point representing the patient’s
final
response will be used
in forming the isopter.
Erasing entries
Repeating points along
the same meridian
11.htp
3/13/03, 3:01 PM
13
Содержание HUMPHREY 720i
Страница 1: ...HUMPHREY FIELD ANALYZER II i series USER S GUIDE Model 720i Model 740i Model 745i Model 750i...
Страница 4: ......
Страница 26: ...1 18 INTRODUCTION INSTRUMENT SETUP REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Страница 116: ...6 8 TEST RELIABILITY REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Страница 208: ...10 20 CUSTOM TESTING REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Страница 284: ...12 14 CARE AND CLEANING REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Страница 327: ......
Страница 328: ......