8-16
FILE FUNCTIONS
REV. B 02/03 PN 51680-1
HUMPHREY FIELD ANALYZER
II -
i
series
Transferring tests from
Transferring results between the HFA I and HFA II is done via serial cable using the
the HFA I to the HFA II
TRANSFER TESTS function.
There are certain limitations to which tests will transfer from the HFA I to the HFA II.
A name
and date of birth must be included on the original HFA I test in order to transfer it
to the HFA II.
The transfer process will not warn you of improperly labeled HFA I tests or test
results which are not eligible to be transferred. The following is a list of restrictions specific to
the test type being transferred:
Threshold
•
Fast Threshold or Master files will not transfer
•
Compared, Averaged or Merged tests will not transfer
•
Tests originally acquired on the HFA II will not transfer
The Macula test: Only the first two threshold values at each point will be displayed and printed.
Note: The Peripheral 30/60-2 test on the HFA I is the Peripheral 60-4 on the HFA II. The test
results will transfer, but the 4 points not found in the 60-4 pattern will not be included.
Screening
All screening tests will transfer from the HFA I to HFA II except:
•
The C-166 Screening test
•
A screening test where the user selected the Central or Peripheral Reference Level
•
Auto Diagnostic tests
Note: The Peripheral 68 test on the HFA I is the P-60 on the HFA II. These test results will
transfer, but the 8 points not found in the P-60 pattern will not be included.
Custom
Custom Arc and Profile tests will not transfer.
Kinetic
Kinetic test results cannot be transferred from the HFA I.
8.htp
3/14/03, 3:18 PM
16
Содержание HUMPHREY 720i
Страница 1: ...HUMPHREY FIELD ANALYZER II i series USER S GUIDE Model 720i Model 740i Model 745i Model 750i...
Страница 4: ......
Страница 26: ...1 18 INTRODUCTION INSTRUMENT SETUP REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Страница 116: ...6 8 TEST RELIABILITY REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Страница 208: ...10 20 CUSTOM TESTING REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Страница 284: ...12 14 CARE AND CLEANING REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Страница 327: ......
Страница 328: ......