CUSTOM TESTING
10-9
REV. B 02/03 PN 51680-1
HUMPHREY FIELD ANALYZER
II -
i
series
4
The Custom screen now
displays each point in the
grid.
5
To input an additional set of Grid points for the same test, repeat steps 1 through 4. There
can be a maximum of 248 points in a test.
When you have completed the entry of Custom points, you need to save the pattern for future
testing. See “Saving Custom Tests” later in this section.
Notes:
1.
Each point in the last grid
you enter will appear as a
small square, not a point.
The previously entered Grid
will turn from squares to
points. This is to make the
most recently programmed
points more noticeable on
the Custom screen. This will
not affect the size of the
points projected during the
test.
2.
If you add a grid that is either too close to a previously entered point or completely
outside of the selected field size, the instrument will reject your selection. Change the
point spacing either by pressing CHANGE OPTIONS and selecting a different spacing
option or by placing the grid in a different location.
3.
If you create a Grid where some points fall outside the selected field size, only the
points within the field will be tested.
10.htp
3/14/03, 2:43 PM
9
Содержание HUMPHREY 720i
Страница 1: ...HUMPHREY FIELD ANALYZER II i series USER S GUIDE Model 720i Model 740i Model 745i Model 750i...
Страница 4: ......
Страница 26: ...1 18 INTRODUCTION INSTRUMENT SETUP REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Страница 116: ...6 8 TEST RELIABILITY REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Страница 208: ...10 20 CUSTOM TESTING REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Страница 284: ...12 14 CARE AND CLEANING REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Страница 327: ......
Страница 328: ......