KINETIC TESTING
REV. B 02/03 PN 51680-1
HUMPHREY FIELD ANALYZER
II -
i
series
4
Press SELECTION COMPLETE. After a slight pause, the HFA II will automatically present
each stimulus to the patient for mapping.
Notes:
1.
The Humphrey Field Analyzer II
does not
automatically monitor the patient’s fixation
during kinetic perimetry. For reliable test results, it is imperative that the operator
constantly monitor fixation with the video eye monitor.
2.
The HFA II will automatically map the blind spot according to the test parameters
chosen. The user cannot initiate each test stimulus individually. When the patient
presses the response button, the HFA II will switch to the next meridian and
immediately start testing.
5
During the test, a small
diamond moves across the
video display. It displays the
progress of the stimulus in the
test bowl.
If it is necessary to stop in the
middle of the blind spot
mapping, touch the STOP
STIMULUS button.
6
When the Blind Spot Map
finishes, the test stops. An
uppercase letter will mark each
location where the patient first
pressed the response button.
Note: Each time a new isopter or test sequence is performed, a new uppercase letter will
be displayed on the test screen. Even if the same stimulus value used for mapping the
blind spot was used for a previous isopter, a different letter will be indicated.
the diamond
the blind spot
11.htp
3/13/03, 3:01 PM
32
Содержание HUMPHREY 720i
Страница 1: ...HUMPHREY FIELD ANALYZER II i series USER S GUIDE Model 720i Model 740i Model 745i Model 750i...
Страница 4: ......
Страница 26: ...1 18 INTRODUCTION INSTRUMENT SETUP REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Страница 116: ...6 8 TEST RELIABILITY REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Страница 208: ...10 20 CUSTOM TESTING REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Страница 284: ...12 14 CARE AND CLEANING REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Страница 327: ......
Страница 328: ......