STATPAC ANALYSIS & PRINTING RESULTS
REV. B 02/03 PN 51680-1
HUMPHREY FIELD ANALYZER
II -
i
series
Three-in-One printout
If you have selected a threshold test pattern or test parameters that do
not
meet the criteria
for STATPAC analysis, results are presented on the Three-in-One printout. The Three-in-One
printout includes a graytone, numeric, and defect depth presentation of the results of a single
test on one page. The numbers which appear outside each quadrant of the numeric grid are
called “quad totals” and represent a summation of the threshold values determined in each
quadrant. These numbers, which even in normals will not be the same for each quadrant, can
be useful in comparing several tests on the same patient over time. This format is available for
10-2, 24-2, 30-2 central field tests, and for the peripheral 60-4 test.
The nasal step threshold test printout differs from other Three-in-One printouts. Only thresh-
old and defect depth data are shown. All graytone data is omitted.
Figure 7.9: Three-in-One Printout
Graytone
Numeric
Defect Depth
7.htp
3/14/03, 3:17 PM
20
Содержание HUMPHREY 720i
Страница 1: ...HUMPHREY FIELD ANALYZER II i series USER S GUIDE Model 720i Model 740i Model 745i Model 750i...
Страница 4: ......
Страница 26: ...1 18 INTRODUCTION INSTRUMENT SETUP REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Страница 116: ...6 8 TEST RELIABILITY REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Страница 208: ...10 20 CUSTOM TESTING REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Страница 284: ...12 14 CARE AND CLEANING REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Страница 327: ......
Страница 328: ......