DATABASE MANAGEMENT
9-19
REV. B 02/03 PN 51680-1
HUMPHREY FIELD ANALYZER
II -
i
series
Information can be restored to the hard disk from the backup magneto-optical disks using the
BACKUP/RESTORE feature. In order to make sure that the information restored to your hard
disk is up-to-date, it is important to make backups of your hard disk frequently.
1.
Start at the File Functions screen, select BACKUP/RESTORE.
2.
Insert the (first) magneto-optical disk into the magneto-optical disk drive.
3.
Choose MAGNETO-OPTICAL as the Source.
4.
Select PROCEED.
5.
You are given the option of replacing the database or merging the database.
Press REPLACE.
6.
A screen will appear showing the progress of the restoration. Should multiple magneto-
optical disks be necessary, follow the directions on the screen. When completed, be sure
to return the disk to a safe place.
7.
Copy to your hard disk any tests stored only to the floppy disk since your last backup.
Note: The option referred to in Step 5 will not be available if the hard disk database has
been disabled following a database failure. The hard disk is disabled as a result of
choosing SAVE TO FLOPPY ONLY.
If you have backed up your entire database onto a series of backup floppy disks, perform the
following steps to restore your database:
1.
Start at the File Functions screen. Select BACKUP/RESTORE.
2.
Choose FLOPPY as the Source. Press PROCEED.
3.
You are given the option of replacing the database or merging the database.
Press REPLACE.
4.
Insert the first backup floppy disk. Press OK.
5.
Insert the remaining disks, as requested.
6.
When complete, return the backup floppy disks to a safe place.
7.
Copy to your hard drive all tests stored to the chronological floppy disk since your last
backup.
Note: The option referred to in Step 3 will not be available if the hard disk database has
been disabled following a database failure. The hard disk is disabled as a result of
choosing SAVE TO FLOPPY ONLY.
Caution: Carl Zeiss Meditec does NOT recommend a floppy disk backup of the
entire database as your only method of data backup. If during restoration one of
the disks is damaged, the entire set of disks becomes unusable.
A floppy disk can have problems similar to a hard disk when it comes to database retrieval.
Occasionally, data cannot be accessed from a floppy disk. Rebuilding the floppy disk database
may fix the problem.
1.
From the Additional Setup screen, press REBUILD FLOPPY DATABASE. A message will
appear asking “Are you sure you want to rebuild the floppy disk database?”. Press OK to
start the rebuild process or CANCEL to return to the Additional Setup screen.
To restore the hard disk
database from magneto-
optical disks
To restore the hard disk
database from backup
floppy disks
Floppy disk failure:
REBUILD FLOPPY DISK
DATABASE
9.htp
3/14/03, 3:20 PM
19
Содержание HUMPHREY 720i
Страница 1: ...HUMPHREY FIELD ANALYZER II i series USER S GUIDE Model 720i Model 740i Model 745i Model 750i...
Страница 4: ......
Страница 26: ...1 18 INTRODUCTION INSTRUMENT SETUP REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Страница 116: ...6 8 TEST RELIABILITY REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Страница 208: ...10 20 CUSTOM TESTING REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Страница 284: ...12 14 CARE AND CLEANING REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Страница 327: ......
Страница 328: ......