FILE FUNCTIONS
8-17
REV. B 02/03 PN 51680-1
HUMPHREY FIELD ANALYZER
II -
i
series
To transfer tests from the
To transfer tests via serial cable:
HFA I to the HFA II - i series
via serial cable
1.
Connect the HFA I and HFA II - i series with an IBM PC compatible serial connector
cable (9 pins on one end and 25 pins on the other). Plug the end of the cable with
nine (9) pins into the Data Transfer port on the HFA II (refer to diagram in Section 1
for port location). Plug the 25 pin connector into the serial port on the back of the
HFA I. Refer to “Additional Components” in Section 1 of this guide for a detailed
illustration of the ports on the HFA II - i series.
2.
Turn both instruments ON.
ON THE HFA II - i SERIES:
3.
From the System Setup screen, select SAVE/TRANSMIT OPTION.
4.
Make sure the TRANSMIT VIA button indicates Serial. Select SERIAL TRANSMIT
SETTINGS.
5.
On the RS-232 Options screen, set BAUD RATE to 9600, PARITY to Even, DATA BITS
to 7, and STOP BITS to 1.
6.
Press PROCEED. The Transmit/Save screen will then appear. Press PROCEED.
7.
Press the
FILE FUNCTIONS
icon on the right of the screen.
8.
Select TRANSFER TESTS.
9.
Select HFA SERIAL CABLE as the Source.
10.
Choose HARD DRIVE or FLOPPY as the Destination, then press PROCEED.
ON THE HFA I:
11.
From the Main Menu, select CONFIGURATION MENU.
12.
Make sure the BAUD RATE is set to 9600 and the PARITY is set to EVEN on the
Configuration Menu screen, then select RETURN (to the Main Menu).
13.
Select DISK FUNCTIONS.
14.
Select TRANSMIT FILES.
15.
Select the disk drive on which the tests you want to transfer are stored.
16.
If you want to transmit a limited number of files, highlight those files and choose
SELECTION COMPLETE. If you want to transmit all the files on the HFA I disk, do not
highlight any files. Just choose SELECTION COMPLETE.
ON THE HFA II - i SERIES:
17.
A pop-up window will update the user on the progress and success of test transfer.
18.
When the “tests transferred” count on the pop-up window has not changed for
several seconds, test transfer is complete. Press CANCEL. If you do not press CANCEL
within about 10 minutes, the HFA II will automatically proceed to the next step.
19.
The transferred tests will then be moved from temporary to permanent storage. Press
OK when the pop-up window informs you that the transmission is complete.
8.htp
3/14/03, 3:18 PM
17
Содержание HUMPHREY 720i
Страница 1: ...HUMPHREY FIELD ANALYZER II i series USER S GUIDE Model 720i Model 740i Model 745i Model 750i...
Страница 4: ......
Страница 26: ...1 18 INTRODUCTION INSTRUMENT SETUP REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Страница 116: ...6 8 TEST RELIABILITY REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Страница 208: ...10 20 CUSTOM TESTING REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Страница 284: ...12 14 CARE AND CLEANING REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Страница 327: ......
Страница 328: ......