PROBLEM
POSSIBLE CAUSE OR SOLUTION
Blind Spot Monitor
Can’t change Blind Spot size
• HFA II uses same stimulus size for testing and Blind Spot check (3)
Can’t find blind spot
• Realign patient (5)
• Verify that proper eye is being tested (5)
• Check to make sure the patient’s non-testing eye is occluded (5)
• Make sure patient is not looking around but fixating on fixation light (5)
• Select FIXATION button and then press RE-TRY TO FIND BLIND SPOT (5)
Blind Spot alarm beep
• Same as “Can’t find blind spot” above
keeps sounding
• Blind Spot may be turned off during test by pressing FIXATION
Gaze Tracking
Gaze Tracking will not initialize
• High power trial lens being used (5)
• Droopy eyelid - ask patient to open eye wider during initialization (5)
• Interfering eyelashes - ask patient to open eye wider during initialization
• Small pupils (5)
• Excessive eye movement or blinking (5)
• Excessively large or dilated pupils (5)
• Dry eyes, cloudy media or dark iris (5)
• Deep-set eyes (5)
Head Tracking
Head Tracking does not work
• Head Tracking not turned on (2)
• Trial lens holder not in Up position (5)
• Gaze Tracking was not initialized at the start of the test (5)
Head Tracking alarm
• Patient’s chin not moving with the chin rest - re-instruct patient (5)
activated too often
• Reset Head Tracking (5)
• Turn off Head Tracking - press TURN OFF HEAD TRACKING
Vertex Monitoring
Vertex Monitoring does not work
• Vertex Monitoring not turned on (2)
• Trial lens holder not in Up position (5)
• Gaze Tracking was not initialized at the start of the test (5)
Vertex Monitoring alarm beep
• Patient moving away from trial lens - reposition patient (5)
goes off too often
• Vertex Monitoring needs to be reset - press RE-INITIALIZE VERTEX (5)
• Turn off Vertex Monitoring - press TURN VERTEX OFF (5)
B-H Appendex.htp
3/14/03, 3:24 PM
28
Содержание HUMPHREY 720i
Страница 1: ...HUMPHREY FIELD ANALYZER II i series USER S GUIDE Model 720i Model 740i Model 745i Model 750i...
Страница 4: ......
Страница 26: ...1 18 INTRODUCTION INSTRUMENT SETUP REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Страница 116: ...6 8 TEST RELIABILITY REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Страница 208: ...10 20 CUSTOM TESTING REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Страница 284: ...12 14 CARE AND CLEANING REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Страница 327: ......
Страница 328: ......