TEST PARAMETERS AND STRATEGIES
4-5
REV. B 02/03 PN 51680-1
HUMPHREY FIELD ANALYZER
II -
i
series
Large Diamond
The Large Diamond is located below the Central
target and is useful for patients with central
scotoma who cannot see either the Central
fixation light or the Small Diamond.
Bottom LED
Some tests have points in the superior visual field
that require a lower fixation light than the central
target. The target used is the Bottom LED of the
Large Diamond. When testing with the Superior 64
or Superior 36 Screening Speciality tests, the
Bottom LED is the default fixation target. It is auto-
matically illuminated at the start of a test.
Fixation
Gaze/Blind Spot
The Blind Spot and Gaze Monitoring system
Monitoring
(model 740i - 750i) are both activated.
Gaze Track
The Gaze Track system automatically measures gaze
(model 740i - 750i) direction at the time of stimulus presentation. Refer
to Section 5: “Gaze Tracking” for more information.
Blind Spot
The test program periodically presents a stimulus
(Heijl-Krakau)
in the patient’s blind spot. If the patient is fixating
well, he or she should not see the blind spot check
stimulus. The Blind Spot check stimulus always
matches the test stimulus size. Refer to Section 6:
“Fixation Losses” for additional information.
Off
Disables Gaze Track and Blind Spot fixation moni-
toring. The operator should monitor fixation with
the video eye monitor.
Blue-Yellow
OFF/
ON
Model 750i and model 745i (optional on model
740i) can perform Blue-Yellow testing. Blue-Yellow
testing uses a Size V blue stimulus presented on a
yellow background. Selecting the Blue-Yellow
option will cause the system to default to these
parameters. See “Blue-Yellow Testing” later in this
section for more information.
Because screening strategies have been designed
and optimized for White-on-White testing, it is not
recommended that screening tests be performed
with the Blue-Yellow testing strategy.
Screening
Parameter
Description
Parameters
Settings
4.htp
3/14/03, 3:13 PM
5
Содержание HUMPHREY 720i
Страница 1: ...HUMPHREY FIELD ANALYZER II i series USER S GUIDE Model 720i Model 740i Model 745i Model 750i...
Страница 4: ......
Страница 26: ...1 18 INTRODUCTION INSTRUMENT SETUP REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Страница 116: ...6 8 TEST RELIABILITY REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Страница 208: ...10 20 CUSTOM TESTING REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Страница 284: ...12 14 CARE AND CLEANING REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Страница 327: ......
Страница 328: ......