10-2
CUSTOM TESTING
REV. B 02/03 PN 51680-1
HUMPHREY FIELD ANALYZER
II -
i
series
This section discusses how to create your own Custom static tests. These tests can be created
as either Screening of Threshold tests. Custom tests are always created for the right eye only.
When the left eye is tested, the HFA II will flip the test pattern to accommodate for the mirrored
physiological differences. Do not attempt to design a Custom test for the left eye.
Note: You may begin testing with either the right or left eye.
Custom tests
must
be created by the user. Once created, they may be saved and then accessed
for testing by selecting the SHOW TEST LIBRARY button from the Main Menu screen. Up to ten
(10) Custom tests can be stored in the Custom Tests screen at any one time.
Note: SITA Standard and SITA Fast are not available for use with Custom tests.
1
Press the
SYSTEM
SETUP
icon from the Main
Menu screen.
2
Choose
ADDITIONAL SETUP.
CREATING
CUSTOM TESTS
Getting to the
custom test screen
10.htp
3/14/03, 2:43 PM
2
Содержание HUMPHREY 720i
Страница 1: ...HUMPHREY FIELD ANALYZER II i series USER S GUIDE Model 720i Model 740i Model 745i Model 750i...
Страница 4: ......
Страница 26: ...1 18 INTRODUCTION INSTRUMENT SETUP REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Страница 116: ...6 8 TEST RELIABILITY REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Страница 208: ...10 20 CUSTOM TESTING REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Страница 284: ...12 14 CARE AND CLEANING REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Страница 327: ......
Страница 328: ......